Magnetic Hotplate-Stirrer with Timer User Manual OptiMag-ST Magnetic Hotplate-Stirrer with Timer Please read the User Manual care
• Set the target stirring speed and begin. • Observe the stirring bar and LCD display. • Set the target temperature and start heating. • Observe
6.1 Working with the PT1000 external temperature sensor Figure 6 The PT1000 external temperature sensor is the manufacturer’s optional accessory. I
7 Stirring The “stirring” function is switched on or off via rotating stirring knob. The speed is set on the knob (100 to 1500 rpm in steps of 10
• Mode of transmission: Fully Duplex. 1 start bit; 7 character bits; 1 parity bit [straight (even)); 1 stop bit • Transmission speed: 9600 bit/s N
• Before using other method for cleaning or decontamination, the user must ascertain with the manufacturer that this method will not harm the ins
13 Specifications Items Digital model Voltage [VAC] 200-240 / 100-120 Frequency [Hz] 50 / 60 Power [W] 550 Stirring point position quantity 1 Ma
Contents Service ... 1 Warranty ...
Preface Welcome to the “(Hotplate) Timer Magnetic Stirrer User Manual”. Users should read this Manual carefully, follow the instructions and proce
1 Safety Instructions Warning! • Read the operating instructions carefully before use. • Ensure that only trained staff work with the instrument.
Do not use damaged components. Safe operation is only guaranteed with the accessories listed in the “Accessories” section. Accessories must be secur
3 Inspection 3.1 Receiving Inspection Unpack the equipment carefully and check for any damage which may have arisen during transport. Please contact
Figure 3 5
Items Descriptions LCD Digital Hotplate Timer Magnetic Stirrer Function switch knob Turning the knob, switches the parameters setting interface, fo
4.2 Display Figure 5 Characters Descriptions Temp & °C Displays temperature when the heating function is switched ON. Stir Displays stirring spe